Novel nanostructured biomaterials: implications for coronary stent thrombosis

Logothetidis, Stergios, Karagkiozaki, Varvara, Karagiannidis, Panagiotis, Kalfagiannis, Nikolaos, Patsalas, Panagiotis, Georgiou, Despoina and Kavatzikidou, Paraskevi (2012) Novel nanostructured biomaterials: implications for coronary stent thrombosis. International Journal of Nanomedicine. pp. 6063-6076. ISSN 1178-2013

Item Type: Article

Abstract

Background: Nanomedicine has the potential to revolutionize medicine and help clinicians to treat cardiovascular disease through the improvement of stents. Advanced nanomaterials and tools for monitoring cell–material interactions will aid in inhibiting stent thrombosis. Although titanium boron nitride (TiBN), titanium diboride, and carbon nanotube (CNT) thin films are emerging materials in the biomaterial field, the effect of their surface properties on platelet adhesion is relatively unexplored.
Objective and methods: In this study, novel nanomaterials made of amorphous carbon, CNTs, titanium diboride, and TiBN were grown by vacuum deposition techniques to assess their role as potential stent coatings. Platelet response towards the nanostructured surfaces of the samples was analyzed in line with their physicochemical properties. As the stent skeleton is formed mainly of stainless steel, this material was used as reference material. Platelet adhesion studies were carried out by atomic force microscopy and scanning electron microscopy observations. A cell viability study was performed to assess the cytocompatibility of all thin film groups for 24 hours with a standard immortalized cell line.
Results: The nanotopographic features of material surface, stoichiometry, and wetting properties were found to be significant factors in dictating platelet behavior and cell viability. The TiBN films with higher nitrogen contents were less thrombogenic compared with the biased carbon films and control. The carbon hybridization in carbon films and hydrophilicity, which were strongly dependent on the deposition process and its parameters, affected the thrombogenicity potential. The hydrophobic CNT materials with high nanoroughness exhibited less hemocompatibility in comparison with the other classes of materials. All the thin film groups exhibited good cytocompatibility, with the surface roughness and surface free energy influencing the viability of cells.

Full text not available from this repository.

More Information

Depositing User: Panagiotis Karagiannidis

Identifiers

Item ID: 7769
Identification Number: https://doi.org/10.2147/IJN.S34320
ISSN: 1178-2013
URI: http://sure.sunderland.ac.uk/id/eprint/7769
Official URL: https://doi.org/10.2147/IJN.S34320

Users with ORCIDS

ORCID for Panagiotis Karagiannidis: ORCID iD orcid.org/0000-0002-2709-4161

Catalogue record

Date Deposited: 21 Sep 2017 15:19
Last Modified: 04 Feb 2021 11:41

Contributors

Author: Panagiotis Karagiannidis ORCID iD
Author: Stergios Logothetidis
Author: Varvara Karagkiozaki
Author: Nikolaos Kalfagiannis
Author: Panagiotis Patsalas
Author: Despoina Georgiou
Author: Paraskevi Kavatzikidou